LEAK-S1 Nondestructive sealing tester

 LEAK-S1 non-destructive sealing tester adopts vacuum attenuation test, which is suitable for non-destructive sealing test and package integrity verification of bottles, tubes, cans, boxes, bags and other products in food, pharmaceutical, medical instruments, electronic components and other industries, as well as non-destructive sealing test of ampoule, Xiling bottles and other samples of tiny holes and leakage.

Online consultation
Consultation hotline:0531-88691870

Product Introduction / Product introduction

Product characteristics

Double pressure test principle, non-destructive testing of products, good repeatability of test results, automatic judgment of test results after the end of the test;

High precision pressure test system, suitable for small leak detection, can also identify large leak samples, and automatically determine qualified and unqualified samples;

Imported pneumatic components, imported double pressure sensor, stable and reliable performance;

Multistage vacuum can be preset for the analysis of sample leakage pressure and minimum leakage;

Multiple test mode, multiple product test mode, can store a variety of standard sample types;

Large LCD touch screen display, one key operation, simple and intuitive operation interface;

Vacuum, test and balance time can be adjusted, and can be stored in the database to ensure the consistency of test conditions;

Test curve real-time display, data intelligent statistics, convenient to quickly view the test results;

With automatic data storage, historical data can be quickly viewed and printed;

The leakage performance of micron leakage holes can be tested, and the data results are traceable.

Equipped with micro printer and USB universal data interface, convenient data output and transfer;

Equipped with GMP authority management system, with user management, data audit tracking and other functions.


Test principle

The host is connected with a vacuum attenuation chamber, and the sample is pumped into the test chamber. There is a pressure difference inside and outside the sample, and the gas enters the vacuum attenuation chamber and the host through the leakage hole. The host uses the pressure sensor and the differential pressure sensor to monitor the vacuum degree change in the test chamber, and determines whether the sample is qualified and the minimum leakage amount of the sample through the vacuum degree change.


Test standard

Reference to Standard ASTMF2338, USP40-1207, Test methods for packaging of sterile medical devices YY/T0681.18 - Part 18: Nondestructive testing of packaging for leaks by vacuum attenuation method


Test procedure

Non-destructive testing method is adopted, which does not damage the test samples, and automatically determines the test results to reduce the influence of human factors. It can detect samples with large leaky holes and non-destructive sealed samples.

Put the sample into the test chamber, and vacuum the test chamber according to the set vacuum degree. If the set vacuum degree is reached within a "certain time", the sample is judged to have no large leakage; otherwise, it is judged to be unqualified.

Set the pressure holding time and critical value. If the vacuum degree drops to the set critical value within the set pressure holding time, it is judged that the sample has no medium size leakage hole; otherwise, it is judged that it is unqualified.

The test time is set. If the vacuum attenuation rate is less than the standard value of the vacuum attenuation rate of the sample during the test time, it is judged that the sample has no corresponding leakage hole above the micron level; otherwise, it is judged that it is unqualified.


Test application

Suitable for the whole sealing integrity test of glass bottles, pipes, cans and boxes;

It is suitable for the whole sealing integrity test of plastic bottles, pipes, cans and boxes.

It is suitable for the whole sealing integrity test of metal bottles, pipes, cans and boxes.

Suitable for sealing integrity verification of ampoule, Xylin, infusion bottles and other materials.


Technical index

Vacuum degree: 0.0 ~ -100KPa

Vacuum sensor accuracy: ±0.25%FS

Detection accuracy of aperture: <3um

Vacuum resolution: 0.1Pa

Test chamber: Sample size and type are customized according to sample specifications

Detection principle: double pressure vacuum attenuation test

Vacuum source: vacuum pump

Dimensions: 490mm(L)×370mm(W)×370mm(H)

Power supply: AC220V50Hz

Net weight main engine: 25kg


Product configuration

Standard configuration: main engine, micro printer, power cord, test chamber, vacuum pump