GC6900P Gas Chromatograph (dual channel)

GC6900P gas chromatograph can be widely used in a variety of materials, gas, odor, residue, cigarette pack and other related indicators detection, detection institutions, research institutes, packaging enterprises, food and drug industries for gas chromatography analysis and control of the appropriate tool.

Online consultation
Consultation hotline:0531-88691870

Product Introduction / Product introduction

Professional technology

Double detector (FID+TCD), double column and double gas path system

Solvent purity detection, determination of water content in hydrophilic solvents

Cigarette strip, box wrapping paper, brand paper organic compound detection

Automatic sampling design is convenient for customers to use flexibly

Industrial microcomputer control

7 "touch screen LCD display test parameters

Two samples were prepared in a constant greenhouse

Two samples were tested at a time

Anti - condensation, back - blowing cleaning function, to avoid cross contamination of the sample

The chromatograph has various test modes of constant temperature, programmed temperature rise and mixed control

16 order linear control temperature rise

Over temperature, off temperature protection

Detector modular design, with expansion function

Automatic zero setting of baseline, power failure protection, parameter storage and call functions

Professional computer reverse control chromatography workstation software support, powerful processing function, optional EPC electronic flow automatic control system


Test principle

Gas chromatograph is a kind of multi-component mixture separation, analysis tool, it is inert gas as mobile phase, using chromatographic column separation technology. When a variety of analytical substances enter the chromatographic column, due to the different distribution coefficient of each component in the chromatographic column, the flow rate of each component in the chromatographic column is different. After a certain column length, the mixed components leave the chromatographic column and enter the detector, which is converted into electrical signals after detection and sent to the data processing workstation, so as to complete the detection and analysis of the measured substances.

Test application

Basic application

Solvent residue

Suitable for all kinds of food, medicine packaging plastic film, paper and plastic composite film, cigarette bag and other materials solvent residue analysis test

Purity analysis

Suitable for all kinds of packaging printing solvent purity analysis

Moisture analysis

It is suitable for solvent moisture content analysis in packaging and printing industry

Automatic sampling

It can realize the automatic experiment process and reduce the human error

Extended application

Vinyl chloride monomer analysis

Suitable for the content analysis of vinyl chloride monomer


Technical index

Sample room A temperature

Room temperature +5℃~200℃; Temperature control accuracy: ±0.1℃

Sample room B temperature

Room temperature +5℃~200℃; Temperature control accuracy: ±0.1℃

Sample gas conveying line temperature

Room temperature +5℃~150℃; Temperature control accuracy ±0.5℃

Sampling switch valve box temperature

Room temperature +5℃~150℃; Temperature control accuracy ±0.5℃

Quantitative loop injection volume

1.0mL (0.5mL, 2mL, 3mL can be selected according to user requirements)

Headspace bottle specifications

300mL (Regular)

Sample analysis station

2

Backblow cleaning

0~60000s

Sampling time control

0~60000s

Headspace equilibrium heating time control

0~60000s

Sample transfer time control

0~60000s

Column chamber temperature

Temperature control accuracy

Better than ±0.1℃

Temperature gradient

Column effective area is not greater than 1%

Temperature deviation

The deviation between the set temperature and the displayed temperature is less than 1 ° C

Temperature deviation

The deviation between the set temperature and the actual temperature is no more than 2%

Program temperature order

Level 5 (user requirements)

Heating rate

0.130/min

Linear program temperature range

Below 150℃ per minute at 30℃;

Lower than 300℃ at 15℃ per minute;

Lower than 350℃ at 10℃ per minute;

Initial temperature and final temperature control time

0600 min

Repeatability of program heating

Not more than 2%

Cooling rate

It takes no more than 15 min to decrease from 300℃ to 50℃

(When the temperature of gasification chamber and detector is 300℃)

Gasification chamber

Temperature control accuracy

±0.1room temperature+15200

more than 200±0.2

Test room

Temperature control accuracy

±0.1room temperature+15200

大于200℃为±0.2℃

Hydrogen flame ion detector (FID)

Detection limit

Not more than 5×10-12g/s(benzene)

noise

Not more than 0.025 mV

drift

Not more than 0.15 mV/h

Thermal conductivity cell detector (TCD)

sensitivity

Not less than 3000 mV·mL/mg(benzene)

noise

Not more than 0.035 mV

drift

Not more than 0.5 mV/h

Overall dimension

595mm (L) × 495 mm (W) × 480 mm (H)

Main engine weight

55 kg



Product configuration

Gas chromatograph host, hydrogen flame ion detector, thermal conductivity detector, water analysis column, capillary analysis column, chromatographic workstation

Note: Air source (high purity nitrogen, high purity hydrogen, dry, oil-free air), pure reagent for analysis (solvent added in printing or compounding process), 1mL glass syringe (with no. 5 needle), computer by user.