Working principle
The pre-treated sample is placed between the upper and lower test chambers, tightly sealed, and then the whole system is vacuumed. When a certain vacuum degree is reached, the lower chamber is closed, some test gas is filled into the high pressure chamber (upper chamber), and the pressure of the upper chamber is stabilized, so that a constant pressure difference is formed on both sides of the sample. Under the action of the pressure difference gradient, the gas can be used for the test. From the high pressure side (upper chamber) to the vacuum side (lower chamber) penetration. The pressure variation of vacuum chamber (lower chamber) is measured accurately, and the barrier parameters of the sample are calculated.
Technical specification
(1) Product characteristics
The vacuum sensor and pressure sensor with resolution of 0.1pa are adopted to ensure the accuracy of the test.
The system has the ability of self-detection to avoid continued testing under the fault state;
Water bath temperature control, automatic control of temperature rise and fall, wide temperature control range, extra high test temperature uniformity;
The equipment adopts the ultimate zero point and sealed pipe system, and adopts the baseline technology to remove systematic errors and ensure the repeatability of test results.
Sample leakage prevention structure design, stable and reliable encapsulation sample seal;
Pressure control ability is accurate, the upper chamber has the function of maintaining pressure, control the stability of the test environment;
Embedded computer system control, the use of quality stability components, instrument stable operation, reliable;
High speed vacuum capacity, reduce vacuum time, degassing thoroughly, ensure test accuracy;
Short test time, improve the test efficiency;
Whole-process monitoring and automatic recording, the test process can be reproduced whole-process;
One-key automatic test, test automatically after setting test parameters, automatically judge the end of the test and save the test results;
Vacuum pump self-control technology, during the test process according to the test needs the equipment can automatically shut down the vacuum pump, saving energy.
(2) Technical parameters
Measuring range: 0.05 ~ 50,000 cc/㎡·24h·0.1MPa
Sample quantity: 1 piece
Vacuum resolution: 0.1Pa
Vacuum: < 20Pa
Temperature control range: 15℃ ~ 90℃
Temperature control accuracy: ±0.1℃
Test pressure: 0.1MPa ~ 0.2MPa
Air source pressure: 0.1MPa ~ 0.8MPa
Test gas: O₂, N₂, CO₂ or other gas
Carrier gas interface: diameter 6
Sample thickness: ≤3mm
Test area: Φ97mm
Power supply: 220VAC50Hz
Weight: 45kg
Product configuration
Standard configuration: main engine, vacuum pump, vacuum bellows, gas hose, disk sampler, seal grease, mouse, keyboard, display
Note: Test air source is prepared by yourself